Knowledge Base
Articles
Viscosity describes how easily a liquid flows. A high-viscosity coating is thick and slow moving; a low-viscosity one is thin and runny. For paints and coatings this single property influences almost every stage of use, from how a product is pumped and dosed in production, to how it atomizes at a spray gun, to how it levels out and resists sagging on a vertical surface. Measuring it accurately, with the right method, is how quality, R&D, and operations experts keep a coating performing the way it was designed to.
Some liquids, like water and solvents, have a viscosity that does not change with how fast they are moved. These are Newtonian. Most paints and coatings are not. Their viscosity changes with shear rate, the speed at which the fluid is deformed, and many are thixotropic, meaning they thin under agitation and recover at rest. This is by design: a paint should be thick in the can so pigment stays suspended, thin under the shear of a brush or spray so it applies evenly, and thick again on the wall so it does not run. Because of this, the shear rate a test applies matters as much as the number it produces, and different methods are built to measure at different shear conditions.
The simplest method is a flow cup. A sample fills a cup with a precision orifice in the base, and the time in seconds for it to drain is recorded, then converted to kinematic viscosity. Flow cups are fast, inexpensive, and ideal on the shop floor for lower-viscosity, near-Newtonian products. Industrial Physics supplies TQC Sheen ISO 2431 viscosity cups and immersion (dip) cups for on-site checks. The two governing standards are ISO 2431 and DIN 53211. Flow cups are less suited to thick or strongly non-Newtonian materials.
The Krebs method is long established for architectural and decorative paints. The TQC Sheen Digital Krebs Viscometer (DV2300) drives a paddle in the sample at a constant 200 rpm and measures the torque, which is proportional to viscosity, reporting the result in Krebs units (KU), centipoise (cP), or grams. It reads across roughly 37 to 141 KU (about 10 to 5,000 cP) with accuracy of 2 percent of full scale, and suits routine mid-range paint QC where results are compared against a familiar KU specification.
When you need dynamic viscosity at high shear, close to how a coating behaves during application, the cone-and-plate method is the reference. The TQC Sheen Cone and Plate Viscometer (CP1) places the sample between a shallow cone and a flat plate and controls the shear rate tightly to 10,000 s-1 (BS and ISO requirements) or 12,000 s-1 (ASTM). These rates are representative of application by roller or brush, so the reading reflects real-world behavior. It measures to ISO 2884, part one of which covers the cone-and-plate geometry at high shear, along with ASTM D4287 and BS 3900-A7. It is the method of choice for R&D and formulation work on non-Newtonian coatings.
For the widest range of fluids, a rotating spindle viscometer measures the torque needed to turn a spindle immersed in the sample. The TQC Sheen Multi Spindle Viscometer conforms to the Brookfield method and is used across paints, coatings, inks, adhesives, and many other liquids, offering a broad measurement range through interchangeable spindles. The related Digital Rotothinner (DV2700) is a rotational instrument that also monitors viscosity continuously as solvents and thinners are added, with manual, max-hold, and timed modes across 0.1 to 350 poise (10 to 35,000 cP).
As a rule of thumb: use a flow cup for a quick, low-cost check of thin, near-Newtonian products on the line; use the Krebs viscometer for mid-range decorative and architectural paints measured in KU; use the cone-and-plate viscometer for dynamic, high-shear measurement of non-Newtonian coatings in R&D and formulation; and use a spindle or Rotothinner instrument for broad-range work or for tracking viscosity during thinning. The right choice is driven by the product, the shear condition you care about, and the standard your customer or specification calls for.
Viscosity is temperature sensitive, so the single most important control is measuring at a defined, stable temperature; a few degrees can shift the result noticeably. Other common errors are air bubbles in the sample, worn or damaged flow cup orifices, inconsistent fill or timing, and comparing results from methods that apply different shear rates. Certified viscosity calibration oils, manufactured to ASTM D2162 and traceable to national standards, are used to verify instruments and keep readings accurate and repeatable.
Viscosity control is quality control. It keeps film build, coverage, gloss, and finish consistent, prevents sagging and running defects, and gives production a fast, objective pass or fail on every batch. Measuring it correctly, to the recognized standards, is part of how Industrial Physics helps quality, R&D, and operations experts protect the integrity of their coatings and their customers’ brands. For more background, see our viscosity testing overview and the guide to types of viscometer.
The TQC Sheen Krebs Viscometer DV2300 is a digital rotational viscometer that measures paint and coating viscosity using the KREBS method - a paddle rotating at a constant 200 rpm in a fixed volume of sample. It converts torque into three simultaneous outputs: Krebs units (KU), centipoise (cP) and grams (g), with automatic unit conversion shown on the LCD display. Used by paints and coatings QC laboratories, paint manufacturers, automotive OEM coating suppliers, printing ink producers, adhesive manufacturers and cosmetics QC labs, the DV2300 supports ASTM D562 Krebs-unit specifications alongside direct viscosity readings for a wide range of Newtonian and near-Newtonian fluids.
The TQC Sheen Viscosity Cup ISO 2431 (Dip/Immersion) is a flow-time viscosity cup compliant with ISO 2431 - in dip/immersion form with a handle for point-of-use paint and coating viscosity checks directly from the container. Available in multiple orifice sizes covering the paint viscosity working range.
Our modern digital viscometer conforms to the Brookfield method and it’s ideal for a wide range of viscosity testing applications.
It depends on the product and how it is applied. Use a flow cup for a quick check of thin products, a Krebs viscometer for mid-range decorative paints in KU, and a cone-and-plate viscometer for dynamic, high-shear measurement of non-Newtonian coatings.
A flow cup measures how long a sample takes to drain through an orifice, giving a fast, relative kinematic reading. A rotational viscometer measures the torque needed to turn a spindle, cone, or paddle in the sample, giving dynamic viscosity and the ability to measure at a controlled shear rate, which matters for non-Newtonian coatings.
The TQC Sheen Cone and Plate Viscometer controls shear to 10,000 s-1 for BS and ISO methods, or 12,000 s-1 for ASTM. These rates represent application by roller or brush, so the measured viscosity reflects how the coating behaves in real use rather than at rest.
Viscosity is strongly temperature dependent, so a small change in sample temperature can shift the result. Reliable testing means measuring at a defined, stable temperature, and comparing results only when they were taken under the same conditions.
Krebs units (KU) are a paint viscosity scale from the Krebs paddle method, in which a paddle turns in the sample at 200 rpm and the torque is converted to KU. The Digital Krebs Viscometer reports KU, centipoise, or grams, and is widely used for architectural and decorative paints.
Viscosity is a fluid's resistance to flow. In paints and coatings it governs how a product brushes, sprays, rolls, levels, and resists sagging, and it is one of the most routinely checked quality parameters on the line and in the lab. A coating that is too thick drags and clogs; one that is too thin…
Key terms Term Definition Viscosity A fluid's resistance to flow. Dynamic viscosity Resistance to flow under an applied shear, measured in pascal seconds (Pa.s), poise (P), or centipoise (cP). Kinematic viscosity Dynamic viscosity divided by density, measured in centistokes (cSt); what a flow cup effectively reports. Shear rate The rate at which a fluid is…
© 2026 Industrial Physics. All rights reserved. Privacy Policy Manage consent