Gas used for medical purposes needs to be tested to the highest standards. Regular analysis of medical gas is required to ensure that the oxygen concentration is NLT 99.0% volume.
USP37 Medical Gas requires a specific, selective and positive identification test using a paramagnetic signal.
Following the appropriate United States Pharmacopeia standards, in this case specifically USP37, requires an analyzer which uses a paramagnetic based technology. The Systech Paramagnetic Oxygen Analyzer PM700 is perfectly suited for this medical gas analysis application.
Paramagnetic oxygen measurement is a method of determining oxygen levels. It uses a sensor to analyze the paramagnetic susceptibility of the oxygen molecule, a physical property which distinguishes oxygen from most other gases. Oxygen is attracted into a strong magnetic field and this non-destructive method of testing can be carried out repeatedly. Find out more in our guide to paramagnetic cells technology.
Oxygen in gases is determined using a paramagnetic analyzer, such as our Paramagnetic Oxygen Analyzer PM700 .
The principle of the method is based on the high paramagnetic sensitivity of the oxygen molecule. Oxygen exerts a strong interaction on magnetic fields, which is measured electronically, amplified and converted to a reading of oxygen concentration.
As the paramagnetic effect of oxygen is linear, the instrument must have a suitable range with a readability of 0.1 percent or better. The measurement of oxygen concentration is dependent upon the pressure and temperature ,so if the analyzer does not automatically compensate for variations, it must be calibrated immediately prior to use.
In healthcare, high purity gas is essential as it can play a lifesaving role in treatments and procedures. The most commonly required medical gas is one with a high oxygen concentration. As medical gases are frequently used to treat patients, they are subject to the same sort of stringent testing standards as medicines. Medical gases can also be used in lab environments, for pharmaceutics and biotechnology research, development and testing.
Oxygen has two unpaired electrons in its molecular orbital and its free electrons move on their axis and create a magnetic field. This means that it is a paramagnetic substance. Compounds or molecules that have only paired electrons in their molecular orbital are diamagnetic substances.
Medical Oxygen contains not less than (NLT) 99.0% oxygen by volume.
Acceptance criteria: NLT 99.0% by volume USP37
Mode: paramagnetic oxygen measurement (see Medical Gases Assay 415)
Identification: the paramagnetic signal exhibited by the sample gas in the Assay confirms the presence of oxygen USP37.
Analysis: determine the percentage by volume of oxygen using a suitable paramagnetic analyzer.
The certified standards called for are listed under Reagents, Indicators and Solutions.
European Pharmacopeia 2.5.27 dictates that oxygen in gases be determined using a paramagnetic analyzer.
Our PM700 Paramagnetic Analyzer offers high performance oxygen analysis by utilizing a magneto-dynamic (dumb bell) transducer. It has a non-depleting sensor, meaning it’s a durable and long-lasting testing machine, and it offers user friendly results.
It may be used for the following applications:
We understand just how important it is to meet those high testing standards when it comes to medical gases and our Systech specialist testing equipment is here to meet your needs. Browse a range of highlighted Systech products and benefit from 30 years of experience and the knowledge of dedicated global experts in gas analysis.
A Question Most Quality Managers Have Never Had to Ask For decades, if you wanted to know how well a beverage can was coated, there was really only one place to look: the scale. Coating weight measurement has been the standard across can manufacturing for so long that it rarely gets questioned. It's practical, it's…
Term Definition Scratch resistance A coating's ability to resist being scratched through by a point or edge under load. Mar resistance Resistance to light surface marks that affect appearance without penetrating the coating. Pencil hardness The hardest pencil grade that does not scratch or mark a coating when pushed across it at 45 degrees. Wolff-Wilborn…
If you're signing off a coating against a customer specification, developing a new formulation, or checking that a production batch has cured properly, scratch resistance testing tells you whether the coating will stand up to everyday contact. This guide covers why scratch resistance matters, the three main test methods, the standards behind them, and how…
© 2026 Industrial Physics. All rights reserved. Privacy Policy Manage consent