No — a melt flow indexer (MFI) does not directly measure intrinsic viscosity. Instead, it measures melt flow rate (MFR) or melt volume rate (MVR), which describe how easily a polymer melts and flows under heat and pressure. Intrinsic viscosity, by contrast, is measured in solution and reflects the molecular weight of the polymer.
While they are different tests, there can be a correlation between MFR and intrinsic viscosity, especially for homogeneous polymer types like PET, PLA, or nylon. Typically:
This relationship can be empirically calibrated. By testing a series of polymer samples with known IVs and measuring their MFRs, you can create a calibration curve.
Here is a simplified example of a typical inverse relationship between MFR and IV (note the log scale on the x-axis):

This allows MFR to be used as a quick proxy for IV — but only within a specific polymer type and grade family. The curve will differ significantly between materials and must be verified with regular solution-based IV measurements.
Yes – many manufacturers use melt flow indexers to monitor relative molecular weight during processing. However, for precise molecular characterization or regulatory reporting, intrinsic viscosity must still be measured using solution viscometry.
Intrinsic viscosity (IV) is a critical parameter for PET (polyethylene terephthalate) manufacturers because it directly relates to the molecular weight of the polymer — which influences how the material performs during processing and in final applications.
The Method B test on the MFR300 includes algorithmic calculation of Apparent Intrinsic Viscosity (using a formula), as well as algorithmic calculation of Shear Stress (Pa), Shear Rate (l/sec) and Viscosity (Pa/sec). In the test setup, there is an “I.V. Corre. Factor” option:

Once a user has identified a reliable IV value for their material, they can run a series of Melt Flow tests on the MFR300 to see how much the Apparent IV value given in the results differs by. The % difference is the figure that needs to be entered as the I.V. Corre. Factor.
This will then allow the user to track the change in IV that occurs as Melt Flow Rate changes for the specific material in question during subsequent testing.
PET is highly hygroscopic, meaning it readily absorbs moisture from the air. Even small amounts of retained moisture will cause hydrolytic degradation during melt processing, lowering molecular weight and artificially reducing the measured Intrinsic Viscosity (IV). To obtain accurate, repeatable IV results, PET must be pre-dried to the manufacturer’s specified moisture content before testing. Regardless of brand, all Melt Flow Indexers use the same underlying algorithmic approach to calculate IV from melt flow data, so the preparation steps and testing methodology remain consistent across instruments.
PET must be thoroughly dried before testing to prevent moisture-related degradation, which can artificially lower the measured Intrinsic Viscosity. Drying should be carried out in accordance with the resin supplier’s recommendations, using suitable equipment to reach the specified moisture level. It’s important to handle dried PET in a way that avoids re-absorption of moisture — typically by transferring it directly to the instrument or storing it in sealed, desiccated conditions until use. Consistent and correct drying practices are essential for obtaining reliable and repeatable results.
If Industrial Physics can assist you further with your melt flow needs, please contact us or browse our melt flow products.
Key terms Term Definition Closed loop control Feeding real-time measurement back to line controls so the process adjusts itself automatically to stay on target. Coating weight / coat weight The amount of coating applied per unit area of coil, the parameter held on target. Wet film thickness (WFT) Coating thickness measured immediately after application, before…
On a coil coating line, applying proper thickness of coating or paint is of paramount significance . Too little coating risks under-protection and quality failures and potential claims; too much application wastes expensive paint and adds cost to every meter of coil. Traditionally, coating thickness is checked by manual or offline sampling, which means the…
Closed Loop Control in Coil Coating with SpecMetrix Closed loop control is establishing an automated process where real-time coating thickness measurements are fed back to the coating line's controls, so the line adjusts itself automatically to dynamically adjust to the target coating weight. Instead of relying on manual control, manual checks or offline sampling, the line responds…
© 2026 Industrial Physics. All rights reserved. Privacy Policy Manage consent